Special custom designed DC UPS & AC UPS which is mounted under a stretcher to power all the lifesaving equipment on the stretcher bridge. The output is backed up by a four cell Lithium Ion Phosphate battery pack.
The AMP-K4910 is a Quad-input device (12VDC / 24VDC / 240VAC and Battery) with dual output (12VDC & 240VAC). It includes 4 x12V isolated Li-FePO4 cell chargers. It also includes control inputs for the 12VDC and 240VAC outputs and LOW VOLTAGE BATTERY disconnection. The battery connection can also be controlled externally.
In Stand Alone Mode (without external controller), the battery will be connected as long as the Battery Voltage is higher than 12.8V (2.9V per cell) and 24VDC or MAINS input are present. It will guarantee that any fluctuation in these two inputs will be smoothed by the battery keeping the outputs (12VDC and 240VAC) constant at any time.